📲Devices
Omron – HEM-9210T
Measurement
This device uses the standard blood pressure measurement protocol. The user can define directly on the device whether the measurements are in mmHg or in kPa.
Type
Unit
systolic_blood_pressure
mmHg or kPa
diastolic_blood_pressure
mmHg or kPa
heart_rate
/min
Example
Here you can download a sample observation made from this device:
Foracare – Diamond Mini
The user can define on the device whether the measurements are in mg/dL or in mmol/L.
Type
Unit
blood_glucose
mg/dL or mmol/L
Example
Here is a sample blood glucose measurement:
Thomson – TGBF1257
The user can define on the device whether the measurements are in kg, lb or in lbs.
Type
Unit
body_weight
kg, lb or lbs
body_fat_percentage
%
total_body_water
%
muscle_mass
%
calorie
cal
example
Thomson – Checkme Pro
This device can be operated without a smartphone nor a tablet during the exams. But the transmission requires the use of Kligo mobile application.
Type of measurement
Type
Unit
L1
uV
ST_amplitude
mV
QT_interval
ms
QRS_duration
ms
QTc
ms
PVC
no unit
heart_rate
/min
spO2
%
PI
%
Interpretations
The CheckMe Pro offers additional information for each component. An interpretation is always present for the L1 component. This interpretation tells whether the ECG Rhythm is regular or not. If the device detects arrhythmia during the exam you can find the interpretation code directly on the specific component.
Component
code
Result
Condition
L1
N
Regular ECG Rhythm
normal
L1
A
Irregular ECG Rhythm
RR irregular
heart_rate
L
Low heart beat
HR < 50/min
heart_rate
H
High heart beat
HR > 100/min
QRS_duration
H
High QRS Value
QRS > 120ms
ST_amplitude
L
Low ST Value
ST < -0.2mV
ST_amplitude
H
High ST Value
ST > 0.2mV
PVC
H
Premature Beat
PVC > 0
spO2
A
Abnormal spO2
spO2
N
Normal spO2
For instance if the exam records a low heart rate, the component will look like the following one:
{
/*...*/,
"components": {
/* other components */
"heart_rate": {
"value": 49,
"unit": "/min",
"interpretation" : "L"
}
}
}
If no interpretation is present on a component block it means that the measurement is within the thresholds defined by the device.
Example
You can download a sample file containing the response of an ecg and oxymeter test done via the CheckMe Pro.
NuvoAir – AirNext
This device requires a smartphone or tablet in order to be operated. Using the Kligo mobile application. Once the spirometry test taken you can retrieve the graph under the flow component.
Type
Unit
FVC
L
FEV1
L
PEF
L/s
flow
L/s
FVC_predicted
L
FEV1_predicted
L
ratio
no unit
Last updated
Was this helpful?